首页 | 本学科首页   官方微博 | 高级检索  
检索        


Quantitation of benzodiazepines in urine, serum, plasma, and meconium by LC-MS-MS
Authors:Marin Stephanie J  Coles Rebecka  Merrell Miles  McMillin Gwendolyn A
Institution:ARUP Institute for Clinical and Experimental Pathology, Salt Lake City, Utah 84108-1221, USA. stephaie.marin@aruplab.com
Abstract:A single method for confirmation and quantitation of a panel of commonly prescribed benzodiazepines and metabolites, alpha-hydroxyalprazolam, alpha-hydroxyethylflurazepam, alpha-hydroxytriazolam, alprazolam, desalkylflurazepam, diazepam, lorazepam, midazolam, nordiazepam, oxazepam, temazepam, clonazepam, and 7-aminoclonazepam, was developed for three specimen types, urine, serum/plasma, and meconium. Quantitation was by liquid chromatography tandem-mass spectrometry (LC-MS-MS) using a Waters Alliance-Quattro Micro system. The instrument was operated in multiple reaction monitoring mode with an electrospray ionization source in positive ionization mode. The method was evaluated for recovery, imprecision, linearity, analytical measurement range, specificity, and carryover. Average recovery and imprecision (within-run, between-run, and total % CV) were within +/- 15% of the target concentrations for urine (10 to 5000 ng/mL) and serum/plasma (10 to 2500 ng/mL) and within +/- 20% for meconium (10 to 5000 ng/g). In all, 205 patient specimens were analyzed, and the results compared to a previous in-house gas chromatography-MS method or LC-MS-MS results from an outside laboratory. Oxazepam glucuronide was evaluated as a hydrolysis control for the urine and meconium specimens.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏