Abstract: | C-reactive protein (CRP) as an acute phase protein is an important diagnostic marker for the presence and course of human processes. Out of the acute phase proteins it is one of those the concentrations increase most rapidly with its sensitivity being superior to other markers of inflammation, such as leukocytosis, erythrocytic sedimentation rate, and fever. This study compared two-point-of-care assays with the standard laboratory method Tina-quant CRP processed on a Hitachi 917: the immunofiltration assay NycoCard CRP Whole Blood and the turbidimetric immunoassay Micros CRP. Both methods are carried in the presence of a patient, by using capillary or venous blood. Seventy-eight blood samples were analyzed first in the standard laboratory routine and then by both rapid test assays. The precision of both assays was determined from the confidence interval. The results were statistically analyzed by arithmetic standard deviation mean method, variation coefficient, Spearman correlation index, Wilcoxon and Bland-Altman tests, and Passing-Bablock regression. NycoCard CRP Whole Blood showed a correlation coefficient of R = 0.9838; the precision had a coefficient of variation of CV = 1.8759% while As compared with Tina-quant CRP had R = 0.9934 and CV = 0.9160%. Both assays indicated the same results as Tina-quant CRP. Both Tina-quant CRP and NycoCard CRP Whole Blood give the best fit for the rapid determination of CRP. |