首页 | 本学科首页   官方微博 | 高级检索  
     


Imaging living mice using a 1-T compact MRI system
Authors:Inoue Yusuke  Nomura Yukihiro  Haishi Tomoyuki  Yoshikawa Kohki  Seki Takahiro  Tsukiyama-Kohara Kyoko  Kai Chieko  Okubo Toshiyuki  Ohtomo Kuni
Affiliation:Department of Radiology, Institute of Medical Science, University of Tokyo, Tokyo, Japan. inoueys-tky@umin.ac.jp
Abstract:PURPOSE: To determine the feasibility of imaging living mice with a 1-T compact MRI system and investigate appropriate imaging techniques for use in routine animal experiments. MATERIALS AND METHODS: An MRI system consisting of a 1-T permanent magnet and compact console was used. Images of the entire trunks of living mice were obtained on the system using a T1-weighted three-dimensional fast low-angle shot (3D FLASH) sequence, and image quality was evaluated in relation to imaging techniques. RESULTS: Restraint of respiratory motion improved the image quality. Decreasing the slice thickness reduced artificial inhomogeneity in signal intensity (SI). Substantial effects of TR and FA on image quality were also demonstrated. With the determined techniques, images covering the entire trunk with a voxel size of 0.26x0.26x0.52 mm were acquired in an acquisition time of five minutes 28 seconds and a total experiment time of <20 minutes, and various organs and subcutaneous tumors were clearly visualized. CONCLUSION: The compact MRI system provides images of living mice with acceptable quality in a reasonable time. Considering its convenience, it appears to be suitable for use in routine mouse experiments.
Keywords:magnetic resonance imaging  mouse  3D FLASH  imaging technique  permanent magnet
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号