容积调强弧形治疗冠矢状面剂量验证的初步探讨
引用本文:石锦平,陈立新,谢秋英,张利文,滕建建. 容积调强弧形治疗冠矢状面剂量验证的初步探讨[J]. 中华放射肿瘤学杂志, 2012, 21(4): 377-379. DOI: 10.3760/cma.j.issn.1004-4221.2012.04.024
作者姓名:石锦平  陈立新  谢秋英  张利文  滕建建
作者单位:528000 佛山,中山大学附属佛山医院广东省佛山市第一人民医院肿瘤中心放疗科(石锦平、谢秋英、张利文、滕建建);510060广州,华南肿瘤学国家重点实验室中山大学肿瘤防治中心放疗科(陈立新)
摘    要:目的 初步探讨容积调强弧形治疗(VMAT)在冠状面和矢状面的剂量验证意义。方法 选取 12例癌症患者的VMAT计划,在多插孔圆柱形体模内插入2个0.125 cm3的电离室探头进行点剂量测量。采用平面二维矩阵电离室及其组合体模,分别在冠状面和矢状面摆位方式下进行γ通过率的剂量验证。结果 在高剂量区域(>最大剂量的80%)和低剂量区域(≤最大剂量的80%)内,点剂量测量与计算结果平均误差分别为1.5%和1.7%。在冠状面和矢状面下二维电离室矩阵中心探头测量的点剂量与计算结果的平均差异为1.7%和1.8%。采用2%/2 mm、3%/3 mm标准的平均通过率分别为93.7%、97.2%。 结论 冠状面和矢状面VMAT剂量验证为临床VMAT计划提供了有意义的数据。

关 键 词:容积调强弧形疗法  剂量验证  质量保证  
收稿时间:2011-05-23

The elementary discussion of volumetric modulated arc therapy using the orthogonal plane dose verification
SHI Jin-ping , CHEN Li-xin , XIE Qiu-ying , ZHANG Li-wen , TENG Jian-jian. The elementary discussion of volumetric modulated arc therapy using the orthogonal plane dose verification[J]. Chinese Journal of Radiation Oncology, 2012, 21(4): 377-379. DOI: 10.3760/cma.j.issn.1004-4221.2012.04.024
Authors:SHI Jin-ping    CHEN Li-xin    XIE Qiu-ying    ZHANG Li-wen    TENG Jian-jian
Affiliation:Department of Radiation Oncology, Cancer Center, First People,s Hospital of Foshan, Affiliated Foshan Hospital of Sun Yat-sen University, Foshan 528000,ChinaCorresponding author:CHEN Li-xin, Department of Radiation Oncology, Cancer Center Sun Yat-sen University, State Key Laboratory of Oncology in South China, Guangzhou 510060,China;Email:chenlx@sysucc.org.cn
Abstract:Objective This study was to explore the feasibility of using the orthogonal plane dose formed by the coronal and sagittal plane to verify the volumetric modulated arc therapy (VMAT) plan. Methods The VMAT plans of 12 patients were included in this study. The orthogonal plane dose formed by the coronal and sagittal plane were measured based on the combination of 2D ionization chamber array and multicube phantom,and the point dose were measured based on a multiple hole cylindrical phantom attached with two 0.125 cm3 ionization chamber probes. Results In the measurement of the point dose, the average error was 1.5% in high dose area (more than 80% of maximum), and 1.7% in low dose area (less than 80% of maximum), respectively. The discrepancy of point dose measurement was 1.3% between the 2D ionization chamber array and the VMAT planning system. In the measurement of the orthogonal plane dose, the pass rate of γ were 93.7% for 2%/2 mm and 97.2% for 3%/3 mm. Conclusion It is reliable for using the orthogonal plane dose formed by the coronal and sagittal plane to verify the VMAT plan.
Keywords:Volumetric modulated arc therapy  Dose verification  Quality assurance