首页 | 本学科首页   官方微博 | 高级检索  
     


Clinical accuracy evaluation of femoral canal preparation using the ROBODOC system
Authors:Shunsaku Nishihara  Nobuhiko Sugano  Takashi Nishii  Hisashi Tanaka  Nobuo Nakamura  Hideki Yoshikawa  Takahiro Ochi
Affiliation:(1) Department of Orthopaedic Surgery, Osaka University Medical School, 2-2 Yamadaoka, Suita, 565-0871, Japan;(2) Department of Radiology, Osaka University Medical School, Osaka, Japan;(3) Center of Arthroplasty, Kyowakai Hospital, Osaka, Japan
Abstract:The purpose of this study was to evaluate the clinical accuracy of femoral canal preparation using postoperative reconstructed computed tomography (CT) images of 75 consecutive total hip arthroplasties performed with a two-pin-based ROBODOC system. Intraoperatively, the robot milled the femoral canal according to the preoperative planning performed with preoperative CT data and the ROBODOC workstation. Postoperative CT data was obtained 1 month postoperatively. Anteroposterior and lateral synthetic radiographs and axial images were reconstructed from CT data on the workstation. The mean difference between the preoperative planning and the postoperative CT images was less than 5% in canal fill, less than 1thinspmm in gap, and less than 1° in the mediolateral and anteroposterior alignment. Clinical femoral canal preparation using the ROBODOC system results in a high degree of accuracy.
Keywords:ROBODOC  ORTHODOC  Clinical accuracy evaluation  Femoral canal preparation
本文献已被 SpringerLink 等数据库收录!