首页 | 本学科首页   官方微博 | 高级检索  
检索        


Verification of an assay for quantification of hepatitis C virus RNA by use of an analyte-specific reagent and two different extraction methods
Authors:Forman Michael S  Valsamakis Alexandra
Institution:Division of Microbiology, The Johns Hopkins Medical Institutions, Meyer B1-193, 600 North Wolfe St., Baltimore, MD 21287-7093, USA.
Abstract:Protocols were designed for quantification and detection of hepatitis C virus (HCV) RNA by the use of an analyte-specific reagent (ASR) (Roche COBAS TaqMan48 CTM48] HCV) after manual and automated RNA extraction. The purposes were to determine (i) assay performance characteristics using manual and automated RNA extraction methods, (ii) whether measurable range and limit of detection (LOD) of the ASR assay were influenced by genotype, and (iii) correlation of quantification by CTM48 HCV ASR and COBAS Monitor HCV v. 2.0. For HCV genotype 1 (Gt1), the lower limits of quantification after manual extraction were slightly lower than those for automated extraction (1.0 versus 1.5 log(10) IU/ml). Results were linear up to the highest concentration tested after extraction by both methods (manual, 6.1 log(10); automated, 6.4 log(10)). Similar results were obtained for Gt2 (1.8 to 6.8 log(10) IU/ml) and Gt3 (1.6 to 6.8 log(10) IU/ml) after automated extraction. The LOD of Gt1 virus was 10 IU/ml after manual extraction and between 25 and 37.5 IU/ml after automated extraction. Results with Gt2 and Gt3 viruses were similar after automated extraction (Gt2, between 25 and 50 IU/ml; Gt3, 25 IU/ml). Variability (intrarun and interrun, at concentrations throughout the range of quantification) was
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号