首页 | 本学科首页   官方微博 | 高级检索  
检索        


Influences of environmental noise level and respiration rate on the accuracy of acoustic respiration rate monitoring
Authors:Shizuha Yabuki  Hiroaki Toyama  Yusuke Takei  Toshihiro Wagatsuma  Hiroshi Yabuki  Masanori Yamauchi
Institution:1.Department of Anesthesiology,Tohoku University Hospital,Sendai,Japan;2.Anesthesiology and Perioperative Medicine,Tohoku University School of Medicine,Sendai,Japan;3.Department of thoracic surgery,Institute of development, aging and cancer,Sendai,Japan
Abstract:We tested the hypothesis that the environmental noise generated by a forced-air warming system reduces the monitoring accuracy of acoustic respiration rate (RRa). Noise levels were adjusted to 45–55, 56–65, 66–75, and 76–85 dB. Healthy participants breathed at set respiration rates (RRset) of 6, 12, and 30/min. Under each noise level at each RRset, the respiration rates by manual counting (RRm) and RRa were recorded. Any appearance of the alarm display on the RRa monitor was also recorded. Each RRm of all participants agreed with each RRset at each noise level. At 45–55 dB noise, the RRa of 13, 17, and 17 participants agreed with RRset of 6, 12, and 30/min, respectively. The RRa of 14, 17, and 16 participants at 56–65 dB noise, agreed with RRset of 6, 12, and 30/min, respectively. At 66–75 dB noise, the RRa of 9, 15, and 16 participants agreed with RRset of 6, 12, and 30/min, respectively. The RRa of one, nine, and nine participants at 76–85 dB noise agreed with RRset of 6, 12, and 30/min, respectively, which was significantly less than the other noise levels (P?<?0.05). Overall, 72.9% of alarm displays highlighted incorrect values of RRa. In a noisy situation involving the operation of a forced-air warming system, the acoustic respiration monitoring should be used carefully especially in patients with a low respiration rate.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号