MR angiography of the carotid arteries: parameters affecting image quality
Authors:Luccichenti Giacomo  Cademartiri Filippo  Lucidi Vincenzo  Marchesi Giuseppe  Ugolotti Ugo  Pavone Paolo
Affiliation:Department of Radiology, University of Parma, Via A. Gramsci 14, 43100 Parma, Italy.
Abstract:RATIONALE AND OBJECTIVES: This study was performed to evaluate the relationship between dose levels of contrast medium and image quality in magnetic resonance (MR) angiography of the carotid arteries with fluoroscopically monitored, manually triggered, elliptically ordered image acquisitions. MATERIALS AND METHODS: Twenty-five patients with clinical indications for angiography of the carotid arteries were examined with MR at 1.5 T by using a fluoroscopically monitored, manually triggered, elliptically ordered pulse sequence with the administration of one of three different volumes of gadolinium-based contrast medium. The signal intensities of the vessel lumen and the surrounding tissues were measured in single partitions at the origin of the common carotid artery, the carotid bifurcation, and the intracranial internal carotid arteries. The contrast-to-noise ratio in these regions of interest also was measured. Maximum intensity projection image quality was appraised for blurring, artifacts, venous enhancement, background suppression, and contrast medium distribution. RESULTS: No artifacts or venous enhancement was observed. The position of the fluoroscopic section affected the distribution of contrast medium along the vessel, as evidenced by the difference between the contrast-to-noise ratio at the origin of the common carotid artery and the ratio at the carotid bifurcation and the intracranial internal carotid arteries (P < .01). The contrast medium dose administered was strongly correlated with image quality (r = 0.90). CONCLUSION: Contrast medium dose is related to image quality in MR angiography of the carotid arteries performed with elliptical ordering, fluoroscopic monitoring, and manual triggering.
Keywords:Arteries, MR   magnetic resonance (MR), contrast enhancement   magnetic resonance (MR), vascular studies
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号