首页 | 本学科首页   官方微博 | 高级检索  
     

智能气囊夹板治疗不稳定型Colles骨折的单中心前瞻性临床研究
引用本文:严培军,刘锌,魏成建. 智能气囊夹板治疗不稳定型Colles骨折的单中心前瞻性临床研究[J]. 南京中医药大学学报, 2023, 39(1): 10-15. DOI: 10.14148/j.issn.1672-0482.2023.0010
作者姓名:严培军  刘锌  魏成建
作者单位:南京中医药大学附属医院, 江苏 南京 210029
基金项目:国家自然科学基金面上项目81973872江苏省中医药管理局项目YB201816
摘    要:  目的  通过单中心的临床随机对照, 进行前瞻性研究, 探究智能气囊夹板相比普通夹板治疗不稳定型Colles骨折的有效性及安全性。  方法  研究纳入157例不稳定型Colles骨折患者, 其中对照组127例使用普通夹板, 治疗组30例使用智能气囊夹板。随访6周, 观察对比2组掌倾角和尺偏角、Lidstrom影像学评分、Gartland-Werley临床功能评分、VAS疼痛评分、夹板调整次数和骨折再复位次数等指标, 并统计2组并发症发生率。  结果  第28、42天时VAS疼痛评分比较差异有统计学意义(P < 0.05);智能气囊夹板再调整次数和骨折再复位次数更少, 差异有统计学意义(P < 0.05)。各时间点2组尺桡骨掌倾角、尺偏角和Lidstrom影像学评分、Gartland-Werley临床功能评分方面无统计学差异(P>0.05)。治疗组患者皮肤和软组织并发症区域面积显著低于对照组(P < 0.05)。  结论  智能气囊夹板在减轻Colles骨折断端疼痛、减少夹板调整次数与骨折再复位次数方面优于普通夹板, 且其压力智能调节显著降低了固定局部的并发症发生率, 值得临床尝试推广使用。 

关 键 词:充气夹板   Colles骨折   临床研究
收稿时间:2022-10-22

Single-center Prospective Clinical Study of Intelligent Pneumatic Splint for Unstable Colles Fractures
Affiliation:The Affiliated Hospital of Nanjing University of Chinese Medicine, Nanjing 210029, China
Abstract:  OBJECTIVE  To investigate the efficacy and safety of intelligent pneumatic splints compared with common splints in the treatment of unstable Colles fractures in a single-center, randomized, controlled, prospective study.  METHODS  A total of 157 patients with unstable Colles fractures were included in the study, 127 of whom were splinted with common splints while 30 of whom were splinted with intelligent pneumatic ones. At 6 weeks of follow-up, the palmar inclination and ulnar deviation angles, Lidstrom imaging score, Gartland-Werley clinical function score, VAS pain score, as well as the number of splint adjustments and fracture resets were compared between the two groups. In addition, the complication rates of both groups were counted.  RESULTS  There was a statistically significant difference in VAS pain scores at 28 and 42 days (P < 0.05). The number of intelligent pneumatic splint readjustments and fracture resets was fewer, and the difference was statistically significant (P < 0.05). However, there was no statistically significant difference in the ulnar and radial palm inclination angle, ulnar deviation angle, Lidstrom imaging score, as well as Gartland-Werley clinical function score between both groups at each follow-up time point (P>0.05). The skin and soft tissue complication area was significantly less in the treatment group than in the control group (P < 0.05).  CONCLUSION  The intelligent pneumatic splint is better than the common splint in reducing the pain of Colles fracture, and reducing the number of splint adjustments and fracture resets. Besides, its intelligent pressure adjustment significantly reduces the incidence of local complications of fixation, which is worthy of clinical application. 
Keywords:
点击此处可从《南京中医药大学学报》浏览原始摘要信息
点击此处可从《南京中医药大学学报》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏