首页 | 本学科首页   官方微博 | 高级检索  
检索        


A novel MRI-compatible brain ventricle phantom for validation of segmentation and volumetry methods
Authors:Khan Amanda F  Drozd John J  Moreland Robert K  Ta Robert M  Borrie Michael J  Bartha Robert;Alzheimer's Disease Neuroimaging Initiative
Institution:Medical Biophysics, Medical Sciences Building, The University of Western Ontario, London, ON, Canada.
Abstract:

Purpose:

To create a standardized, MRI‐compatible, life‐sized phantom of the brain ventricles to evaluate ventricle segmentation methods using T1‐weighted MRI. An objective phantom is needed to test the many different segmentation programs currently used to measure ventricle volumes in patients with Alzheimer's disease.

Materials and Methods:

A ventricle model was constructed from polycarbonate using a digital mesh of the ventricles created from the 3 Tesla (T) MRI of a subject with Alzheimer's disease. The ventricle was placed in a brain mold and surrounded with material composed of 2% agar in water, 0.01% NaCl and 0.0375 mM gadopentetate dimeglumine to match the signal intensity properties of brain tissue in 3T T1‐weighted MRI. The 3T T1‐weighted images of the phantom were acquired and ventricle segmentation software was used to measure ventricle volume.

Results:

The images acquired of the phantom successfully replicated in vivo signal intensity differences between the ventricle and surrounding tissue in T1‐weighted images and were robust to segmentation. The ventricle volume was quantified to 99% accuracy at 1‐mm voxel size.

Conclusion:

The phantom represents a simple, realistic and objective method to test the accuracy of lateral ventricle segmentation methods and we project it can be extended to other anatomical structures. J. Magn. Reson. Imaging 2012;36:476–482. © 2012 Wiley Periodicals, Inc.
Keywords:3T  brain phantom  MRI  ventricle  software validation  segmentation
本文献已被 PubMed 等数据库收录!