A comparison between size of the occluder device and two-dimensional transoesophageal echocardiographic sizing of the ostium secundum atrial septal defect
Authors:Alimohammad Hajizeinali  Mohammad Alidoosti  Hakimeh Sadeghian  Arezoo Zoroufian  Mehrnaz Rezvanfard  Marat A Volman
Abstract:

Objectives

Transcatheter closure of a secundum atrial septal defect (ASD II) has become an effective alternative for surgical treatment. In this study we evaluated the correlation between the two-dimensional transoesophageal echocardiographic (2D TEE) sizing of ASDs and the actual diameter of occluders in patients undergoing device closure.

Methods

The records of 54 patients who underwent transcatheter ASD closure were reviewed. ASD characteristics and maximum defect diameter were evaluated using preprocedure 2D TEE images. Appropriate device size was determined by the balloon sizing method, which measures the balloon occlusive diameter (BOD) via TEE and fluoroscopy. ASD closure was performed under continuous TEE monitoring using the Amplatzer occluder in all patients.

Results

The mean of the TEE-derived maximum defect diameter was significantly lower than the mean of the BOD (17.8 ± 4.5 vs 22.1 ± 5.1 mm; p < 0.001) and the mean size of the implanted occluder device (17.8 ± 4.5 vs 23.3 ± 5.1 mm; p < 0.001). However, a good correlation was found between the TEE-derived defect size and the BOD (BOD = 0.898 × TEE defect size + 6.212, R = 0.824; p < 0.001) and between the TEE measurement and the final size of the implanted Amplatzer (device size = 0.928 × TEE defect size + 6.853, R = 0.822; p < 0.001).

Conclusions

2D TEE may provide a good equation to predict the BOD or the size of the occluder device; however, further studies are needed to investigate whether it is feasible to perform transcatheter ASD occlusion without balloon sizing.
Keywords:atrial septal defect (ASD)   occluder device   2D TEE   balloon sizing