Measuring the arterial input function with gradient echo sequences.
Authors:Matthias J P van Osch  Evert-jan P A Vonken  Max A Viergever  Jeroen van der Grond  Chris J G Bakker
Institution:Department of Radiology, Image Sciences Institute, University Medical Center Utrecht, The Netherlands. thijs@isi.uu.nl
Abstract:The measurement of the arterial input function by use of gradient echo sequences was investigated by in vitro and in vivo experiments. First, calibration curves representing the influence of the concentration of Gd-DTPA on both the phase and the amplitude of the MR signal were measured in human blood by means of a slow-infusion experiment. The results showed a linear increase in the phase velocity and a quadratic increase in DeltaR(*) (2) as a function of the Gd-DTPA concentration. Next, the resultant calibration curves were incorporated in a partial volume correction algorithm for the arterial input function determination. The algorithm was tested in a phantom experiment and was found to substantially improve the accuracy of the concentration measurement. Finally, the reproducibility of the arterial input function measurement was estimated in 16 patients by considering the input function of the left and the right sides as replicate measurements. This in vivo study showed that the reproducibility of the arterial input function determination using gradient echo sequences is improved by employing a partial volume correction algorithm based on the calibration curve for the contrast agent used.
Keywords:arterial input function  dynamic susceptibility contrast MRI (DSC‐MRI)  cerebral perfusion  contrast agent