Department of Cardiology, Kyoto Second Red Cross Hospital, 355-5 Haruobi, Kamanza-dori Marutamachi-agaru, Kamigyo, Kyoto, Japan. mhyogo@gaea.ocn.ne.jp
Abstract:
We experienced an entrapped rotablator burr that could not be retrieved even by deep seating of the guiding catheter. We successfully retrieved the burr by balloon inflation after the tapered tip of a Conquest wire managed to penetrate the quite hard plaque and pass through the outer lumen of the burr.