显像时间与活性对PET/CT图像质量的影响
引用本文:冯彦林,贺小红,黄克敏,余丰文,刘德军,袁建伟,袁白虹,苏少弟.显像时间与活性对PET/CT图像质量的影响[J].中国医学影像技术,2005,21(1):127-130.
作者姓名:冯彦林  贺小红  黄克敏  余丰文  刘德军  袁建伟  袁白虹  苏少弟
作者单位:广东省佛山市第一人民医院PET/CT中心,广东,佛山,528000
摘    要:目的探讨显像时的活性水平与显像时间对PET/CT图像质量的影响.方法分别在Jaszczak模型内高活性短采集时间、推荐活性和采集时间以及低活性长采集时间下,以全身显像方式采集发射数据;采用X线CT作衰减校正(CTAC)和全身显像方式进行图像重建.相同患者按5.18 MBq/kg给药,分别于给药50 min和90 min后采集数据,采用CTAC和全身显像方式进行图像重建.以152.026 MBq、每床位采集3 min 30 s对1例74 kg女性志愿者进行显像.结果三种显像条件所得图像都可清晰分辨模型中直径为6.4 mm的冷、热区,且均匀区域未见伪影.给药50 min后开始采集所得患者图像清晰,但有时病灶难以与正常组织区分;给药90 min后开始采集所得患者图像清晰,对比度良好,能进一步明确病灶轮廓与数量.低活性较长采集时间也能获得较好的图像质量.结论通过对活性和显像时间的平衡,明显降低了二者对PET/CT图像质量的影响.给药与开始采集数据的时间间隔最好在90 min以上.

关 键 词:体层摄影术  发射型计算机  图像质量  质量保证  质量控制  脱氧葡萄糖
文章编号:1003-3289(2005)01-0127-04
收稿时间:2004/9/10 0:00:00
修稿时间:2004年9月10日

Impact of injected activity and scanning time to the image quality of PET/CT
FENG Yan-lin,HE Xiao-hong,HUANG Ke-min,YU Feng-wen,LIU De-jun,YUAN Jian-wei,YUAN Bai-hong and SU Shao-di.Impact of injected activity and scanning time to the image quality of PET/CT[J].Chinese Journal of Medical Imaging Technology,2005,21(1):127-130.
Authors:FENG Yan-lin  HE Xiao-hong  HUANG Ke-min  YU Feng-wen  LIU De-jun  YUAN Jian-wei  YUAN Bai-hong and SU Shao-di
Institution:PET/CT Center, Department of Nuclear Medicine, the First People's Hospital of Foshan, Foshan 528000, China;PET/CT Center, Department of Nuclear Medicine, the First People's Hospital of Foshan, Foshan 528000, China;PET/CT Center, Department of Nuclear Medicine, the First People's Hospital of Foshan, Foshan 528000, China;PET/CT Center, Department of Nuclear Medicine, the First People's Hospital of Foshan, Foshan 528000, China;PET/CT Center, Department of Nuclear Medicine, the First People's Hospital of Foshan, Foshan 528000, China;PET/CT Center, Department of Nuclear Medicine, the First People's Hospital of Foshan, Foshan 528000, China;PET/CT Center, Department of Nuclear Medicine, the First People's Hospital of Foshan, Foshan 528000, China;PET/CT Center, Department of Nuclear Medicine, the First People's Hospital of Foshan, Foshan 528000, China
Abstract:
Keywords:Tomography  emission-computed  Image quality  Quality assurance  Quality control  Deoxyglucose