首页 | 本学科首页   官方微博 | 高级检索  
     


Quality control procedures for flow cytometric applications in the hematology laboratory
Authors:J.-L. D’hautcourt
Affiliation:1. Laboratoire de Cytométrie, H?pital de Warquignies, 27, rue des Chaufours, B-7300, Boussu, Belgique
Abstract:Clinical diagnosis is one of the areas in which flow cytometry (FCM) has gained wide popularity and FCM now plays a crucial role in several aspects of medical hematology. It has progressively replaced many traditional laboratory tests due to its greater accuracy, sensitivity and rapidity. Unfortunately, among the very large number of its potential applications, only a minority of flow cytometric protocols have been standardized. Numerous factors are responsible for variation in analytical conditions and may affect results obtained by FCM. All these variables can be schematically divided into three major groups: factors related to the biological samples, immunological and accessory reagent factors and factors associated with the use of instruments. The quality control program must monitor and evaluate all aspects of the procedure. This includes the following main aspects: 1) performance of the flow cytometer, 2) specimen collection, transportation and maintenance of its integrity, 3) reagents, particularly monoclonal antibodies and 4) sample measurement, data acquisition and their interpretation. Procedures described here are designed to assess all the settings which affect the reliability, reproducibility and sensitivity of the cytometer in order to ensure identical conditions on a daily basis.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号