首页 | 本学科首页   官方微博 | 高级检索  
     

高频磁刺激仪的降低功耗设计
引用本文:李松,殷涛,安好,刘志朋. 高频磁刺激仪的降低功耗设计[J]. 医疗卫生装备, 2009, 30(6): 4-6,9
作者姓名:李松  殷涛  安好  刘志朋
作者单位:中国医学科学院,北京协和医学院,生物医学工程研究所,天津,300192
基金项目:天津市重点科技攻关项目,国家科技部升级改造项目 
摘    要:目的:经颅磁刺激技术在精神、心理疾病治疗方面具有优势,但由于达到神经电位阈值所需的刺激能量较大,对电源功耗的要求限制了磁刺激的高频应用,设计一种磁刺激电路降低功率消耗是必要的。方法:设计了由单片机控制的两组开关.分剐控制对无极性电容的正反向充放电.通过电压检测等手段控制充放电时序;对系统软硬件的设计进行了初步实验验证,实验参数包括采用电容为25.1μF,电感为0.29mH,回路电阻为1.87n,电源输出电压最高为100V.输出功率最高为200W.刺激频率为1~100Hz,放电时间250μs。结果:磁刺激电路使用正反双向充放电电路与单向充放电电路进行对比试验,结果表明其能量消耗降低了11%。结论:设计的正反双向充放电方法可以有效降低高频磁刺激仪的功耗.提高磁刺激仪器能量利用率。

关 键 词:磁刺激  RCL电路  正反向充电  功率消耗

Design of Lower-Power Expenditure in High-frequency Stimulation
LI Song,YIN Tao,AN Hao,LIU Zhi-peng. Design of Lower-Power Expenditure in High-frequency Stimulation[J]. Chinese Medical Equipment Journal, 2009, 30(6): 4-6,9
Authors:LI Song  YIN Tao  AN Hao  LIU Zhi-peng
Affiliation:(Institute of Biomedical Engineering Chinese Academy of Medical Sciences, Peking Union Medical College,Tianjin 300192, China)
Abstract:Objective To develop a device to reduce the power expenditure, because magnetic stimulating technology takes advantages of clinical therapies in spiritual and psychological diseases, therefore the electromagnetic energy of each pulse should be large enough to exceed the threshold of nerve action potentials, the required power of magnetic stimulation limited the application in high frequency. Methods A circuit controlled by microcomputer is proposed to ontrol two cluster switches charging or discharging the non-polar capacitor in positive and negative direction. Voltage detection is used to control the actions of charging and discharging the capacitor. Validating experiments are carried out to test the system function of the hardware and software. The parameter of the experiments is set as: L of 0.29 mH; C of 25.1 μF; R of 1.87Ω. Power output is set as 100 V and 200 W, frequency is 100 Hz and discharge time is 250 μs. Results By comparing the bidirectional charging and discharging stimulating process with the single charging and discharging stimulating process, the energy expenditure is saved 11%. C, lladutllaa This design has lower-power expenditure effectively, which goes further to improve the utilization of energy.
Keywords:magnetic stimulation  RCL circuit  charge bidirectionaly  power expenditure
本文献已被 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号