首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison of the sheath delivery system versus bare stenting for coronary stent implantation
Authors:Shunji Kasaoka  Rirei Son  Mahmoud Eslami  Clare Pierman  Jesse Currier  Lawrence A. Yeatman  Jonathan M. Tobis
Abstract:Outside the United States, Palmaz-Schatz coronary stents are implanted by hand-crimping the stent to a high pressure balloon without the use of a protective sheath. This lowers the delivery profile, increases the ease of deployment, and ensures that the postdilatation balloon is centered on the stent. To assess this bare stenting technique, 209 patients were retrospectively analyzed: 92 patients (107 lesions) with the sheath protected stent delivery system (SDS) and 117 patients (150 lesions) with the bare stent approach. The number of balloons used per lesion in the bare stent group was significantly less than in the SDS group (1.9 ± 0.6 vs. 3.8 ± 1.2, P < 0.0001). In addition, the procedure time in the bare stent group was significantly shorter than in the SDS group (106 ± 55 vs. 134 ± 60 min, P = 0.001). There was no difference in frequency of adverse events or stent displacement during the procedure. The bare stenting technique decreases the procedure time, reduces the number of balloons used, and is as safe as the SDS approach. Cathet. Cardiovasc. Diagn. 43:386–394, 1998. © 1998 Wiley-Liss, Inc.
Keywords:coronary stents  stent delivery system  bare stenting
设为首页 | 免责声明 | 关于勤云 | 加入收藏