首页 | 本学科首页   官方微博 | 高级检索  
检索        


Port site electrosurgical (diathermy) burns during surgical laparoscopy
Authors:P D Willson  J D van der Walt  D Moxon  J Rogers
Institution:(1) Academic Surgical Unit, Royal London Hospital, London, E1 1BB, United Kingdom, GB;(2) Department of Histopathology, Royal London Hospital, London, E1 1BB, United Kingdom, GB
Abstract:Background: Direct and capacitive coupling of diathermy current have been reported as causes of occult injury during surgical laparoscopy. Methods: In order to determine the incidence of electrosurgical injury adjacent to metal and plastic cannulas, skin biopsies at 19 port sites used for monopolar electrosurgery were analyzed for coagulative necrosis. Prior to surgery the cannulas were randomized to either metal or plastic. Results: Coagulative necrosis was observed at nine electrosurgery port sites compared to only one control (χ2= 4.872; df= 1; 0.05 > p > 0.02). Plastic cannulas afforded no greater protection from skin burns than metal cannulas. Conclusions: Burns may be the result of direct or capacitive coupling to metal cannulas or capacitive coupling to the skin edge across plastic cannulas. The potential exists for burns to other tissues also in close proximity to a cannula used for electrosurgery. Received: 12 August 1996/Accepted 26 November 1996
Keywords:: Electrosurgery —  Surgical diathermy —  Capacitive coupling —  Direct coupling —  Patient burns
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏