首页 | 本学科首页   官方微博 | 高级检索  
     


Pitfalls in the application of distal femur plates for fractures
Authors:Collinge Cory A  Gardner Michael J  Crist Brett D
Affiliation:Department of Orthopaedic Trauma, Harris Methodist Fort Worth Hospital and John Peter Smith Orthopaedic Surgery Residency Program, Fort Worth, TX 76104, USA. ccollinge@msn.com
Abstract:Despite design features intended to aid the surgeon in restoring proper alignment, malunion and implant-related problems are relatively common after a distal femur fracture treated with plate fixation. This article presents case examples of these problems followed by a discussion of the relevant distal femoral anatomy, design features of modern locked distal femur plating systems, and technical points necessary to avoid malunion and implant-related problems when using these devices.
Keywords:
本文献已被 PubMed 等数据库收录!