IOL Master与Pentacam对角膜曲率及前房深度测量的比较
引用本文:宋超,孟觉天. IOL Master与Pentacam对角膜曲率及前房深度测量的比较[J]. 临床眼科杂志, 2012, 20(1): 5-7
作者姓名:宋超  孟觉天
作者单位:1. 徐州医学院,221002
2. 徐州市眼病防治研究所,徐州市第一人民医院,221002
摘    要:目的比较IOL Master与Pentacam测量角膜曲率和前房深度的差异。方法纳入166例(332只眼)正常眼,分别采用IOL Master、Pentacam测量角膜曲率(K1、K2)和前房深度(ACD)与角膜中央厚度(CCT)之和。应用配对t检验比较两种仪器测量平均角膜曲率(K)[K=(K1+K2)/2]及ACD与CCT之和的差异及相关性。结果 IOL Master和Pentacam测得的K值分别为(43.49±1.41)D(,43.39±1.40)D,差异有统计学意义(P〈0.001),测得的ACD与CCT之和分别为(3.78±0.25)mm(、3.80±0.31)mm,差异没有统计学意义(P=0.316)。两种仪器的测量结果均有显著相关关系,相关系数分别为r=0.991,P〈0.001;r=0.721,P〈0.001。结论 IOLMaster和Pentacam均可较准确的测量正常眼的角膜曲率与前方深度,两种仪器在临床应用上可相互替代。

关 键 词:角膜曲率  前房深度  IOL  Master  Pentacam

Comparison of keratometry and anterior chamber depth measurements taken with the IOL Master and Pentacam
SONG Chao , MENG Jue-tian. Comparison of keratometry and anterior chamber depth measurements taken with the IOL Master and Pentacam[J]. Journal of Clinical Ophthalmology, 2012, 20(1): 5-7
Authors:SONG Chao    MENG Jue-tian
Affiliation:.Department of Ophthalmology,the First People’s Hospital of Xuzhou,Xuzhou Eye Institution,Jiangsu,221002,China
Abstract:Objective To compare the difference of keratometry and anterior chamber depth(ACD) measured by the IOL Master and Pentacam.Methods A total of 166 cases(332 eyes) was involved.Keratometry(K1,K2)、anterior chamber depth(ACD) and corneal center thickness(CCT) were taken with the IOL Master and Pentacam respectively.Paired t tests were performed to compare the difference and correlation of mean keratometry K=(K1+K2)/2 and the sum of ACD and CCT between the two instruments.Results The mean K value was(43.49±1.41)D for IOL Master,and(43.39±1.40)D for Pentacam.There was statistically significant difference in K value(P<0.001).The sum of ACD and CCT was(3.78±0.25)mm for IOL Master,(3.80±0.31)mm for Pentacam.NO statistically significant difference(P=0.316) was found.Relative high correlation of results measured by IOL Master and Pentacam were revealed,r=0.991,P<0.001;r=0.721,P<0.001,repectively.Conclusion IOL Master and Pentacam can measured keratometry and ACD of normal eyes accurately.The two instruments can be interchangeable in clinical application.
Keywords:Keratometry  Anterior chamber depth  IOL Master  Pentacam