Echocardiographic evaluation of iatrogenic atrial septal defect after catheter-based mitral valve clip insertion |
| |
Authors: | Saitoh Takeji Izumo Masaki Furugen Azusa Tanaka Jun Miyata-Fukuoka Yoko Gurudevan Swaminatha V Tolstrup Kirsten Siegel Robert J Kar Saibal Shiota Takahiro |
| |
Affiliation: | Cedars-Sinai Heart Institute, Cedars-Sinai Medical Center, and UCLA, Los Angeles, California, USA. |
| |
Abstract: | The geometries and sizes of persistent iatrogenic atrial septal defects (IASDs) after transseptal puncture during catheter-based mitral valve clip insertion (MVCI) have not been detailed. In this study, 11 IASDs were investigated in 10 patients who underwent MVCI using a guide catheter (24Fr proximally and 22Fr at the atrial septum). The diameters of the long and short axes and the area at maximum and minimum during a cardiac cycle were measured after MVCI using real-time 3-dimensional (RT3D) transesophageal echocardiography (TEE). A circular shape was assumed on 2-dimensional TEE, resulting in an area calculation of π × (dimension/2)(2). The anatomic geometries of IASDs were visualized in a 3-dimensional en face view of the atrial septum. Furthermore, 1 month after MVCI, IASDs were evaluated using echocardiography. The IASDs had a variety of irregular geometries. The mean long-axis diameter was 1.0 ± 0.24 cm, the mean short-axis diameter was 0.51 ± 0.22 cm, and the mean area was 0.40 ± 0.24 cm(2) on RT3D TEE. The diameters and area changed significantly between the maximal and minimal values during the cardiac cycle. Importantly, 2-dimensional TEE underestimated the maximal diameters of IASDs (0.54 ± 0.17 vs 1.0 ± 0.24 cm by RT3D TEE, p <0.01) and the maximal areas of IASDs (0.25 ± 0.15 vs 0.40 ± 0.23 cm(2) by RT3D TEE, p <0.05). One month after MVCI, the smallest and the second smallest IASDs had closed, and the other 9 remained open. In conclusion, RT3D TEE is useful to assess the irregular geometries of IASDs created during MVCI. |
| |
Keywords: | |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|