Performance evaluation of the Artus hepatitis C virus QS-RGQ assay |
| |
Authors: | Paba Pierpaolo Fabeni Lavinia Perno Carlo Federico Ciotti Marco |
| |
Affiliation: | Laboratory of Molecular Virology, Foundation Polyclinic Tor Vergata, Viale Oxford, 81-00133 Rome, Italy |
| |
Abstract: | Accurate determination of hepatitis C virus RNA level is essential for evaluating the response to antiviral therapy, to determine the duration of treatment, and to predict treatment outcome. Currently, two real-time based polymerase chain reaction assays are used widely to monitor the hepatitis C RNA level: the Abbott RealTime HCV assay and the Cobas Taqman HCV assay. Recently, a third assay has become commercially available: the Artus HCV QS-RGQ assay, which uses the QIAsymphony SP/AS platform for sample preparation and PCR-setup, and the Rotor-Gene Q for amplification and detection. In this study, the performance of the Artus HCV QS-RGQ assay was tested on 105 plasma samples and compared to that of the Cobas Taqman HCV assay. Linear regression analysis showed a good agreement between the two assays. A slightly better sensitivity was observed with the Cobas Taqman assay, while higher hepatitis C viral RNA levels were measured by the Artus HCV QS-RGQ assay in samples positive for hepatitis C genotypes 4. Taken together, the data suggest that the Artus HCV QS-RGQ assay is useful in a diagnostic setting. The combination with the versatile QIAsymphony SP/AS system may represent a major advantage for clinical virological laboratories aiming at optimizing their workflow. |
| |
Keywords: | HCV real-time PCR HCV viral load |