首页 | 本学科首页   官方微博 | 高级检索  
     


Comparative study of four methods for quantifying doppler ultrasound waveforms from the femoral artery
Authors:K.W. Johnston  M. Kassam  J. Koers  R.S.C. Cobbold  D. MacHattie
Affiliation:Institute of Biomedical Engineering and the Department of Surgery, University of Toronto, Toronto, Ontario, Canada M5S 1A4
Abstract:Currently, there is no agreement as to the best method for quantifying Doppler ultrasound recordings from peripheral arteries in order to detect occlusive disease. The four methods assessed in this study are: the pulsatility index, height-width index, path length index, and a Laplace transform function index. Recordings of the Doppler ultrasound spectral waveforms from the common femoral artery of 232 limbs were digitized to obtain the maximum velocity waveforms. This data was analyzed and the various indices were computed and then compared with the arteriographic grades. The effect of distal disease was also examined. The diagnostic accuracy of each index was determined from receiver operating characteristic curves. We concluded that all four indices were capable of detecting significant aortoiliac disease with approximately equal diagnostic accuracy of 90–95% but that pulsatility index had the advantages of simplicity and ease of calculation.
Keywords:Doppler ultrasound  Aortoiliac disease  Pulsatility index  Laplace transform indices
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号