Intradevice interaction in a dual chamber implantable cardioverter defibrillator preventing ventricular tachyarrhythmia detection
Authors:Shivkumar K  Feliciano Z  Boyle N G  Wiener I
Affiliation:Division of Cardiology, UCLA Medical Center, UCLA School of Medicine, Los Angeles, California, USA.
Abstract:Failure to detect ventricular tachycardia and/or ventricular fibrillation by implantable cardioverter defibrillators (ICDs) is a rare but serious problem. We report a case of failure to detect an episode of induced ventricular tachycardia by a dual chamber ICD, due to abbreviation of ventricular detection window secondary to programmed pacing parameters and a rate-smoothing algorithm. In this patient, the intradevice interaction was corrected by programming rate-smoothing off. This report highlights the potentially lethal consequences of critical timing relationships among the pacing function, arrhythmia detection, and the characteristics of the arrhythmia when using a modern dual chamber ICD. Physicians responsible for patients with ICDs must be aware of such interactions.
Keywords:implantable cardioverter defibrillator    arrhythmia detection