首页 | 本学科首页   官方微博 | 高级检索  
     


Design and biomechanical study of slide-poking external fixator for hip fracture
Authors:Hua-Biao Chen  Hong-Bo Wu  Min Chen  Yu-Liang Huang
Affiliation:Department of Orthopaedic Surgery, Affiliated Huizhou Central People Hospital, Sun Yat-Sen University and Guangdong Medical University, Huizhou City, Guangdong Province, People’s Republic of China
Abstract:BackgroundFemoral head collapse and coxa vara lead to internal fixator failure in elderly patients with hip fracture. External fixator application is an optimal choice; however, the existing methods have many disadvantages.MethodsType 31-A1.3 hip fracture models were developed in nine pairs of 1-year-old fresh bovine corpse femur specimens. Each left femur specimen was fixed by a dynamic hip screw (control group), and each right femur specimen was fixed by the slide-poking external fixator (experimental group). Vertical loading and torsion tests were then performed in both groups.ResultsIn the vertical loading experiment, a 1000-N load was implemented. The mean vertical downward displacement of the femoral head in the experimental and control groups was 1.49322 ± 0.116280 and 2.13656 ± 0.166374 mm, respectively. In the torsion experiment, when the torsion was increased to 10.0 Nm, the mean torsion angle in the experimental and control groups was 7.9733° ± 1.65704° and 15.4889° ± 0.73228°, respectively. The slide-poking external fixator was significantly more resistant to compression and rotation than the dynamic hip screw.ConclusionThe slide-poking external fixator for hip fractures that was designed and developed in this study can provide sufficient stability to resist compression and rotation in hip fractures.
Keywords:Design   hip fracture   external fixator   dynamic hip screw   biomechanics   stability
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号