Highly sensitive liquid chromatography-tandem mass spectrometry method for quantification of a new bone anabolic agent,TAK-778, in human serum |
| |
Authors: | Teshima Koichiro Kondo Takahiro Maeda Chie Oda Tsuneo Hagimoto Toshiaki Tsukuda Ryoichi Yoshimura Yoshinobu |
| |
Institution: | Drug Analysis and Pharmacokinetics Research Laboratories, Pharmaceutical Research Division, Takeda Chemical Industries Ltd., 2-17-85 Juso-Honmachi, Yodogawa-ku, Osaka 532-8686, Japan. teshima_kouichirou@takeda.co.jp |
| |
Abstract: | A liquid chromatography-tandem mass spectrometric (LC-MS-MS) method for the highly sensitive determination of a new bone-anabolic agent, TAK-778 in human serum was developed. The internal standard (I.S.) used was deuterated TAK-778. TAK-778 and I.S. were extracted from serum samples with diethyl ether at neutral pH. A turbo ion spray interface was used as the ion source of LC-MS-MS, and the analysis was performed in the selected reaction monitoring mode. The lower limit of quantification was 0.02 ng/ml when 0.4 ml of serum was used, and the standard curve was linear in the range of 0.02-10 ng/ml. The method was precise; the intra- and inter-day precision of the method was not more than 17.9%. The accuracy of the method was good with the deviations between added and calculated concentration of TAK-778 being typically within 9.0%. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|