Clinical evaluation of a wireless ECG sensor system for arrhythmia diagnostic purposes
Authors:Rune Fensli  Torstein Gundersen  Tormod Snaprud  Ole Hejlesen
Institution:1. University of Agder, Faculty of Engineering and Science, Grimstad, Norway;2. Sørlandet Hospital, HF, Medical Department, Arendal, Norway;3. Aalborg University, Department of Health Science and Technology, Denmark
Abstract:In a clinical study, a novel wireless electrocardiogram (ECG) recorder has been evaluated with regard to its ability to perform arrhythmia diagnostics. As the ECG recorder will detect a “non-standard” ECG signal, it has been necessary to compare those signals to “standard” ECG recording signals in order to evaluate the arrhythmia detection ability of the new system. Simultaneous recording of ECG signals from both the new wireless ECG recorder and a conventional Holter recorder was compared by two independent cardiology specialists with regard to signal quality for performing arrhythmia diagnosis. In addition, calculated R–R intervals from the two systems were correlated. A total number of 16 patients participated in the study. It can be considered that recorded ECG signals obtained from the wireless ECG system had an acceptable quality for arrhythmia diagnosis. Some of the patients used the wireless sensor while doing physical sport activities, and the quality of the recorded ECG signals made it possible to perform arrhythmia diagnostics even under such conditions. Consequently, this makes possible improvements in correlating arrhythmias to physical activities.
Keywords:Evaluation studies  Ambulatory electrocardiography  Holter monitoring  Wireless ECG sensor  Clinical trials
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号