Trans-scleral tactile tonometry: An instrumented approach
Authors:Péter P Polyvás  Gholam Peyman  Eniko T Enikov
Institution:1. Advanced Micro & Nano Systems Laboratory, Department of Aerospace and Mechanical Engineering, University of Arizona, 1130 N. Mountain Ave., Tucson, AZ 85721, United States;2. Department of Basic Medical Sciences, University of Arizona College of Medicine – Phoenix, 425 N. Fifth Street, Phoenix, AZ 85004-2157, United States
Abstract:This article describes a feasibility study of a novel trans-scleral tonometer based on the use of an instrumented form of digital palpation tonometry. Similar to manual digital palpation tonometery, trans-scleral tonometer utilizes two force probes offset by a fixed distance. Force indentation data from these probes have been shown to correlate with the intraocular pressure (IOP) of the eye. Enucleated porcine eyes were used to experimentally validate the approach. The observed hysteresis in the force data was analyzed using an analytical model that accounts for the outflow of the aqueous humor. The predictions of the model indicate that the primary reason behind the observed hysteresis is stress relaxation (accommodation) in the visco-elastic corneo-scleral shell. Experimental data from eye distention and indentation tests were then used to infer the conditions under which the novel tonometer would be expected to have an accuracy of ±1 mmHg. Analysis of the data shows that indentation rates should be kept below 0.5 mm/s for a pressure range of 10–35 mmHg. Two commonly used pressure control protocols were tested in an effort to ensure accurate IOP values during the palpation tests. Due to the large increase of IOP during digital palpation, the trans-scleral (intra-vitreous) pressurization was found to be inadequate, leading to clogging of the line by the displaced vitreous. No such problems were identified when the eye was pressurized through the cornea and into the anterior chamber. Force data from multiple palpation experiments are used to generate calibration curves for a two-probe conceptual tonometer. The calibration showed that a 10 mN of force variation corresponds to 1 mmHg of IOP change. A possible implementation using a contoured facial mask is also presented.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号