18F-FDG PET/CT心肌代谢显像受检者的周围辐射剂量水平
引用本文:何鑫雨,皮颖斐,曹雪,武云云,姜晓燕,贠明凯,张晓丽,丁库克.18F-FDG PET/CT心肌代谢显像受检者的周围辐射剂量水平[J].中国辐射卫生,2022,31(2):192-196.
作者姓名:何鑫雨  皮颖斐  曹雪  武云云  姜晓燕  贠明凯  张晓丽  丁库克
作者单位:1. 中国疾病预防控制中心辐射防护与核安全医学所放射生态学研究室,北京 100088;2. 首都医科大学附属北京安贞医院核医学科,北京 100029
基金项目:国家自然科学基金(31770907)
摘    要:目的 探索PET/CT心肌代谢显像受检者周围辐射剂量水平,为保证周围人员辐射安全提供数据支持。方法 选取33例行18F-FDG PET/CT心肌代谢显像患者,在受检者胸部等高位置按照不同体位方向、时间和距离处对其周围剂量当量率进行测量,探索受检者周围剂量当量率的分布规律。结果 在相同测量时间和距离下,受检者身体两侧的周围剂量当量率相比正、背面更低。受检者注射18F-FDG显像剂后,距离体表1 m处正面的周围剂量当量率为13~21 μSv/h,检查结束后降至5~14 μSv/h,平均降低46%。受检者周围剂量当量率随距离(10~300 cm)的增加呈幂函数趋势降低,其幂指数均值为?1.2。结论 心肌代谢显像受检者注射药物后周围辐射水平较高,周围剂量当量率随时间和距离增加快速降低,建议心肌代谢显像受检者检查当天避免与他人长时间、近距离接触。

关 键 词:核医学  辐射防护  周围剂量率  心肌代谢显像  
收稿时间:2021-09-04

Ambient radiation of patients undergoing 18F-FDG PET/CT myocardial metabolism imaging
HE Xinyu,PI Yingfei,CAO Xue,WU Yunyun,JIANG Xiaoyan,YUN Mingkai,ZHANG Xiaoli,DING Kuke.Ambient radiation of patients undergoing 18F-FDG PET/CT myocardial metabolism imaging[J].Chinese Journal of Radiological Health,2022,31(2):192-196.
Authors:HE Xinyu  PI Yingfei  CAO Xue  WU Yunyun  JIANG Xiaoyan  YUN Mingkai  ZHANG Xiaoli  DING Kuke
Affiliation:1. Department of Radioecology National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088 China;2. Department of Nuclear Medicine, Beijing AnzhenHospital, Capital Medical University, Beijing 100029 China
Abstract:Objective To study the ambient radiation of patients undergoing 18F-FDG PET/CT myocardial metabolism imaging, and to provide data for protection of surrounding people against radiation.Methods A total of 33 patients were selected for 18F-FDG PET/CT myocardial metabolism imaging. Dose equivalent rate was measured at the height of patient chest, in different directions, at different time points and at different distances, to investigate the distribution of ambient radiation of the patient.Results At the same time point and distance, the dose equivalent rates at the left and right sides of the patient were lower than the front and back sides. The dose equivalent rate at 1 m in front of the patient was 13-21 μSv/h after 18F-FDG injection, and decreased to 5-14 μSv/h after PET/CT imaging, with a mean decrease of 46%. The ambient dose equivalent rate decreased exponentially with distance (10~300 cm), and the mean power was -1.2.Conclusion The ambient radiation of patients undergoing 18F-FDG PET/CT myocardial metabolism imaging was high after 18F-FDG injection, and the ambient dose equivalent rate decreased rapidly with time and distance. Our results suggest that patients undergoing myocardial metabolism imaging should avoid prolonged and close contact with other people on the day of examination.
Keywords:Nuclear medicine  Radiation protection  Ambient dose equivalent rate  Myocardial metabolism imaging  