首页 | 本学科首页   官方微博 | 高级检索  
检索        


Early Clinical Experience with a Minute Ventilation Sensor DDDR Pacemaker
Authors:GABRIEL VANERIO  SHAILESH PATEL  ELIZABETH CHING  RICHARD G TROHMAN  BRUCE L WILKOFF  LON W CASTLE  JAMES D MALONEY  TONY W SIMMONS
Institution:Pacemaker and Electophysiology Section, Department of Cardiology, The Cleveland Clinic Foundation, Cleveland, Ohio
Abstract:The new DDDR pacemaker META DDDR utilizes a minute veritilation sensor based on transthoracic impedance measurements. The sensor determines the metabolic indicated interval, the atrioventricular (AV) delay and the postventricular atrial refractory period (PVARP). The baseline PVARP must be carefully selected to define nonphysiological tachycardias. If a Pwave falls within the PVARP the pacemaker will automatically switch to the VVIR mode. This behavior prevents tracking of paroxysmal atrial tachyarrhythmias (PAT). Twenty-eight patients with sinus node dysfunction (n = 20), AV junction ablation (n = 5), complete or intermittent AV block (n = 3); who received a META DDDR pacemaker were studied. The mean age was 65 ± 13 years. Results: mode switching (reversion) to VVIR was observed in 57% of the patients. Forty-two percent had episodes of mode switching to VVIR during a stress test four related to PAT, and seven to sinus tachycardia. Fifty percent had episodes of mode switching to VVIR during a 24-hour Holter, four related to PAT, three to retrograde P wave sensing, and two to sinus tachycardia. At the last follow-up, 20 of the 26 patients initially programmed to the DDDR mode remained in the DDDR mode, while five were reprogrammed to the DDD and one to the VVIR mode. Mode switching has a high sensitivity but a low specificity for PAT. It appears to be a useful approach to prevent rapid tracking of atrial tachyarrhythmias. Careful PVARP programming is critical to appropriate reversion behavior, but further modifications of the algorithm are needed to improve its performance.
Keywords:rate responsive pacing  DDDR  minute ventilation  atrial arrhythmias
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号