首页 | 本学科首页   官方微博 | 高级检索  
     


Evaluation of three automated systems for susceptibility testing of enterobacteria containing qnrB, qnrS, and/or aac(6')-Ib-cr
Authors:Calvo Jorge  Cano María Eliecer  Pitart Cristina  Marco Francesc  Rodríguez-Martínez José Manuel  Pascual Alvaro  Martínez-Martínez Luis
Affiliation:Service of Microbiology, University Hospital Marqués de Valdecilla, Av. Valdecilla s/n, 39008 Santander, Spain. jcalvo@humv.es.
Abstract:The accuracy of the MicroScan WalkAway, BD Phoenix, and Vitek-2 systems for susceptibility testing of quinolones and aminoglycosides against 68 enterobacteria containing qnrB, qnrS, and/or aac(6 ')-Ib-cr was evaluated using reference microdilution. Overall, one very major error (0.09%), 6 major errors (0.52%), and 45 minor errors (3.89%) were noted.
Keywords:
本文献已被 PubMed 等数据库收录!