首页 | 本学科首页   官方微博 | 高级检索  
检索        


Energy dependence (75 kVp to 18 MV) of radiochromic films assessed using a real-time optical dosimeter
Authors:Rink Alexandra  Vitkin I Alex  Jaffray David A
Institution:Princess Margaret Hospital/Ontario Cancer Institute, Department of Medical Biophysics, University of Toronto, Toronto, ON M5G 2M9 Canada.
Abstract:The response of radiochromic film, GafChromic EBT, was investigated for dependence on x-ray beam energy using a previously reported real-time optical readout approach. X-ray beams of energy from 75 kVp to 18 MV were employed. The dose-induced change in optical density for the EBT film was compared to values obtained for GafChromic HS and MD-55 films, exposed under the same conditions. All responses were normalized to that obtained for 60Co irradiation. While change in optical density for 1 Gy of applied dose as measured with HS and MD-55 films decreased by approximately 40% at low energies, the mean change in optical density of EBT film remained within 3% of that in the 60Co beam over the entire energy range.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏