Differential pulse cathodic voltammetric determination of floctafenine and metopimazine |
| |
Authors: | Gazy Azza A Hassan Ekram M Abdel-Hay Mohamed H Belal Tarek S |
| |
Affiliation: | Pharmaceutical Analytical Chemistry Department, Faculty of Pharmacy, University of Alexandria, Elmessalah 21521, Alexandria, Egypt. azzagazy@yahoo.com |
| |
Abstract: | A simple, rapid and sensitive voltammetric method for the determination of floctafenine (FFN) and metopimazine (MPZ) was developed. Well-defined cathodic waves were obtained for both drugs in Britton-Robinson buffer pH 9.0 using the differential-pulse mode at the hanging mercury drop electrode (HMDE). The current-concentration relationship was found to be linear over the ranges 0.4-3.6 and 0.4-2.4 microg ml(-1) for FFN and MPZ, respectively. The quantification of the two drugs in their pharmaceutical formulations was carried out using the proposed voltammetric method and compared with spectrophotometric analysis data. The mechanisms of the electrode reactions for the two drugs were proposed. |
| |
Keywords: | Floctafenine Metopimazine Differential-pulse Voltammetric determination HMDE |
本文献已被 ScienceDirect PubMed 等数据库收录! |