景深延长型人工晶状体Symfony ZXR00的视觉质量研究
引用本文:段崇楠 崔蓓 张微 赵明 杨宏宇 贾洪真. 景深延长型人工晶状体Symfony ZXR00的视觉质量研究[J]. 眼科, 2022, 31(5): 355-359
作者姓名:段崇楠 崔蓓 张微 赵明 杨宏宇 贾洪真
作者单位:1.首都医科大学附属北京同仁医院 北京同仁眼科中心 眼科学与视觉科学北京市重点实验室 100730;;2. 解放军总医院第三医学中心眼科医学部派驻第六医学中心眼科,北京 100048
摘    要:目的 评价Symfony ZXR00植入术后患者的视觉质量。设计前瞻性非随机对照研究。研究对象2021年6月至2022年1月于解放军总医院第六医学中心行白内障超声乳化摘除联合IOL植入术的患者42例(42眼),其中植入Symfony ZXR00IOL者18例(18眼),单焦点IOL A1-UV 24例(24眼)。方法比较两组术后3个月视力、对比敏感度、离焦曲线,并对光学干扰现象、脱镜率和视物满意度进行调查和分析。主要指标视力(LogMar)、对比敏感度、离焦曲线、光学干扰现象、脱镜率和视物满意度。结果术后3个月,Symfony组的裸眼远视力(-0.04±0.08)、裸眼中视力(-0.03±0.08)、裸眼近视力(0.28±0.16)、矫正近视力(-0.03±0.05)、矫正远视力下的中视力(-0.04±0.07)、矫正远视力下的近视力(0.27±0.13)显著优于单焦点A1-UV组(0.03±0.10、0.32±0.09、0.39±0.17、-0.01±0.02、0.30±0.11、0.48±0.19),P均<0.05;矫正远视力差异无统计学意义。Symfony组在-2.50 D...

关 键 词:对比敏感度  人工晶状体
收稿时间:2022-04-28

Clinical study of an extended depth-of-focus intraocular lens Symfony ZXR00
Affiliation:1 Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Key Laboratory of Ophthalmology and Visual Science, Beijing 100730, China; 2 Department of Ophthalmology, the Sixth Medical Center stationed by the Senior Department of Ophthalmology of the Third Medical Center of Chinese PLA General Hospital, Beijing 100048, China
Abstract:Objective To evaluate visual quality of patients implanted with the extended depth-of-focus IOL Symfony ZXR00 after cataract surgery. Design Prospective nonrandomized controlled comparative study. Participants The study included 42 patients (42 eyes) who had cataract phacoemulsification surgery in the Sixth Medical Center of Chinese PLA Gereral Hospital from June 2021 to January 2022, wherein Symfony ZXR00 IOLs were implanted in 18 cases (18 eyes), and single focus IOLs A1-UV in 24 cases (24 eyes). Methods Visual acuity (VA), defocus curves and contrast sensitivity (CS) were compared 3 months postoperatively between the two groups. Optical interference, spectacle independence and visual satisfaction were also investigated and analyzed. Main Outcome Measures VA (LogMar), CS, defocus curves, optical interference, spectacle independence and visual satisfaction. Results After 3 months, uncorrected distant VA (UDVA)(-0.04±0.08)、uncorrected intermediate VA (UIVA)(-0.03±0.08)、uncorrected near VA (UNVA)(0.28±0.16)、corrected near VA (CNVA)(-0.03±0.05)、distance-corrected intermediate VA (DCIVA)(-0.04±0.07) and distance-corrected near VA (DCNVA)(0.27±0.13)were better in the Symfony ZXR00 group than in the A1-UV group (0.03±0.10, 0.32±0.09, 0.39±0.17, -0.01±0.02, 0.30±0.11, 0.48±0.19), all  P<0.05. CDVA was similar in the two groups. The Symfony ZXR00 still provided visual acuity greater than 0.3 for a defocus level of -2.50 D. Contrast sensitivity was similar in the two groups at any spatial frequency. The patients in the Symfony group had no serious optical interference, and the spectacle independence rate was 77.8%, which was significantly higher than 0% in the A1-UV group (χ2=28.000, P=0.000). There was no significant difference in the satisfaction rate for distance vision between the two groups. The patient satisfaction rates for intermediate vision in the Symfony group was 100%, which was significantly higher than 20.8% in the A1-UV group (χ2=26.022, P=0.000).For near vision, the satisfaction rates were 77.8% and 0% in the two groups, respectively (χ2=28.000, P=0.000). Conclusion Symfony ZXR00 can provide excellent distance, intermediate and functional near visual acuity, with no significant decline in contrast sensitivity and close to monofocal IOL.   (Ophthalmol CHN, 2022, 31: 355-359)
Keywords:contrast sensitivity  intraocular lens  