人工心瓣在脉动循环模拟系统上的能量损耗的研究
引用本文:李时峰 Blacd,MM. 人工心瓣在脉动循环模拟系统上的能量损耗的研究[J]. 生物医学工程学杂志, 1995, 12(2): 120-125
作者姓名:李时峰 Blacd  MM
作者单位:华西医科大学生物医学工程研究室
摘    要:本文对常用的不同尺寸的5种人工心瓣在脉循环系统的主动脉瓣位,在不同的心率,心输邮量和主动脉平均压下进行能量损耗(Energylosses)的评价。人工心瓣的能量损耗据所计录的左心室压,主动脉压,前反向的跨瓣流量而计算。结果表明,跨瓣压关(ΔP)主输量(CO)的增加而增加,随心率(HR)和心瓣尺寸(VS)的增加而减小;跨瓣反流量(Reg)随VS的增加而增加,且除笼球瓣外,Reg随CO的增加而减少;主

关 键 词:人工瓣膜 脉动循环系统 模拟系统 能量损耗

In Vitro Study of Energy Losses of Heart Valve Prostheses in Aortic Position
Li Shifeng, Chen Huaiqin ,Xue Zhennan. In Vitro Study of Energy Losses of Heart Valve Prostheses in Aortic Position[J]. Journal of biomedical engineering, 1995, 12(2): 120-125
Authors:Li Shifeng   Chen Huaiqin   Xue Zhennan
Abstract:The performance of five types of different size (VS) of heart valve prostheses in the aortic position is assessed based on the in vitro energy losses of valves. A pulse duplicator system was used to test valves under different cardiac output (CO),heart rate (HR)and mean aortic pressure (AP). The energy losses (EL)of valves were calculated from the recorded ventricular and aortic pressure,as well as forward and reverse flows. The results show that transvalvular pressure difference increases with increasing CO,and decreases with increasing HR and VS ; transvalvular regurgitation increases with increasing VS,and decreases with increasing CO except that of StarrEdwards Silastic ball valve (SE) ;AP has litter effect on transvalvular pressure difference and regurgitation. However, the total energy loss (ELtot )of valves decreases with increasing HR,AP and VS,and the ELtot of tissue valves and Stars-Edwards valve increases with increasing CO,but Bjork -Shiley Monostrut and St Jude Medical valves have a higher ELtot at 2. 0 L/m and 8. 0 L/m of CO than at 4. 0 L/m and 6. 0 L/m of CO. In any condition ,for the tissue valve,Sorin-Pericarbon (SP)pericardial valve has a lower ELtot than Carpentier-Edwards (CE) porcine valve does;for the mechanical valve,the SE has a higher ELtot than Bjork-Shiley Monostrut (BS)and St Jude Medical (SJ)do except at 2. 0 L/m of CO and 100 bpm of HR.
Keywords:Heart valve prosthesis Pulse duplicator system Transvalvular pressure difference Regurgitation Energy loss  