A new approach to the noninvasive measurement of cardiac output using an annular array Doppler technique--I. Theoretical considerations and ultrasonic fields
Authors:J M Evans  R Skidmore  N P Luckman  P N Wells
Institution:Department of Medical Physics, Bristol General Hospital, England.
Abstract:This paper describes the development of a Doppler flowmeter capable of measuring blood volume flow rate without the need to measure the vessel lumen area or beam-vessel angle. It requires the production of a uniform wide ultrasound beam to encompass the whole vessel and thus to produce a Doppler spectrum which corresponds to all the flowing blood, and a narrow reference beam placed within the lumen to compensate for various unknown quantities, such as tissue attenuation. The general definition of volume flow rate is described and applied to a new flowmeter, which allows an absolute value of volume flow rate to be measured independently of vessel size, beam-vessel angle, and tissue attenuation. By electronically apodising an annular array transducer in transmission and reception, a uniform wide beam and a narrow reference reception beam are produced. Theory to predict these beam patterns is developed and a computer simulation is made. The ultrasonic fields obtained from an annular array transducer in water are compared with the theoretical fields.