201Tl定量门控心肌灌注体层显像与99mTc-红细胞门控心血池显像测定左心室射血分数的对比研究
引用本文:袁建伟,冯彦林,贺小红,余丰文,刘德军,温广华,黄克敏. 201Tl定量门控心肌灌注体层显像与99mTc-红细胞门控心血池显像测定左心室射血分数的对比研究[J]. 国际放射医学核医学杂志, 2008, 32(2)
作者姓名:袁建伟  冯彦林  贺小红  余丰文  刘德军  温广华  黄克敏
作者单位:广东省佛山市第一人民医院核医学科,528000
摘    要:目的 探讨201Tl定量门控心肌灌注体层显像与99mTc-红细胞门控心血池显像测量左心室射血分数(LVEF)的相关性.方法 72例受检者接受201Tl静息门控心肌灌注体层显像,用AUTOQUANT 4.21软件测量LVEF,并与24 h内的静息99mTc-红细胞平衡法门控心血池显像结果进行比较.结果 ①门控心肌灌注体层显像与门控心血池显像测量LVEF值的结果呈明显正相关(r=0.554,P=-0.000),两种方法无统计学差别(t=1.194,P>0.05).②不同疾病组之间两种测量方法无统计学差异(P值均大于0.05).③门控心肌灌注体层显像及门控心血池显像测量的LVEF值分别为(64.68±10.77)%和(62.46±8.99)%,门控心肌灌注体层显像测量的LVEF值要比门控心血池显像高出3.55%.结论 201Tl门控心肌灌注体层显像与99mTc-红细胞门控心血池显像测量LVEF值的相关性好且结果准确,但门控心肌灌注体层显像的LVEF测量值要稍高于门控心血池显像.

关 键 词:门控血池显像  门控心肌灌注显像  铊放射性同位素  体层摄影术,发射型计算机,单光子

Comparison study of left ventricular ejection fraction measured by 201Tl gated myocardial imaging and 99mTc-red blood cell gated blood-pool imaging
YUAN Jian-wei,FENG Yan-lin,HE Xiao-hong,YU Feng-wen,LIU De-jun,WEN Guang-hua,HUANG Ke-min. Comparison study of left ventricular ejection fraction measured by 201Tl gated myocardial imaging and 99mTc-red blood cell gated blood-pool imaging[J]. International Journal of Radiation Medicine and Nuclear Medicine, 2008, 32(2)
Authors:YUAN Jian-wei  FENG Yan-lin  HE Xiao-hong  YU Feng-wen  LIU De-jun  WEN Guang-hua  HUANG Ke-min
Abstract:Objective To explore the correlation of automatic quantification of left ventricular ejection fraction (LVEF) from gated myocardial perfusion SPECT with 201Tl and rest equilibrium gated bloodpool imaging with 99mTc-red blood cell.Methods Rest gated myocardial perfusion SPECT with 201Tlwas performed on 72 cases.AUTOQUANT 4.21 software was used to measure LVEF.Rest equilibrium gated blood-pool imaging was also performed within 24 h.The values of LVEF measured by two methods were gated myocardial perfusion SPECT and rest equilibrium gated blood-pool imaging(r=0.554,P=0.000).There methods were(64.68±10.77)%and(62.46±8.99)%.The values of LVEF measured by gated myocardial perfusion SPECT was higher 3.55%than rest equilibrium gated blood-pool imaging.Conclusion There was good correlation between the LVEF measured with gated myocardial perfusion SPECT by AUTOQUANT 4.21 software and gated blood-pool imaging.The LVEF can be accurately evaluated using gated myocardial perfusion SPECT with 201Tl.The value of LVEF measured by gated myocardial perfusion SPECT was higher than rest equilibrium gated blood-pool imaging.
Keywords:Gated blood-pool imaging  Gated myocardial imaging  Thallium radioisotopes  Tomography,emission-computed,single photon