Spatial resolution requirements for digital chest radiographs: an ROC study of observer performance in selected cases
Authors:Lams  PM; Cocklin  ML
Abstract:Thirty-eight selected clinical radiographs were digitized and displayed on a 1,024-line monitor at pixel sizes of 1.6, 0.8, 0.4, and 0.2 mm. Eighteen experienced radiologists assessed the radiographs and digital images, which included 12 examples of abnormal solitary nodular density, ten examples of septal lines, and 16 controls, six of which showed diffuse lung abnormalities. For each level of spatial resolution and for film reading, observers gave their decision confidence on a sliding scale of probability. Receiver operating characteristic curves were generated from these data. It was found that while spatial resolution requirements for solitary nodules were not critical for pixel sizes at or below 0.8 mm, the requirement for septal lines was likely to be 0.4 mm (1.25 line pairs/mm).
Keywords:
点击此处可从《Radiology》浏览原始摘要信息
点击此处可从《Radiology》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏