首页 | 本学科首页   官方微博 | 高级检索  
     


Dysfunction of an atrial septal defect occluder 8 years after implantation
Authors:Ten Freyhaus Henrik  Rosenkranz Stephan  Südkamp Michael  Höpp Hans-Wilhelm
Affiliation:Klinik III für Innere Medizin, Universit?t zu K?ln, Cologne, Germany.
Abstract:Catheter interventional treatment of atrial septal defect (ASD) is widely accepted. The ASD occluder system (ASDOS) is no longer a widely used device nowadays. However, it is implanted in a substantial number of patients. We report a case of severe left-to-right shunt 8 years after catheter interventional closure of an ASD with an ASDOS device. The shunt was due to a membrane perforation, while the arms of the device were not dislocated. Microscopy, microbiology, and histology could not establish a proper explanation for the dysfunction; so long-term follow-up investigation may be required in patients with an implanted ASDOS device.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号