标准化西格玛性能验证图在常规生化项目性能评价中的应用
作者姓名:陈真  陈滔  黄珍  熊伟
作者单位:吉首市人民医院检验科;湘西州民族中医院检验科
摘    要:目的 应用标准化西格玛验证图评估实验室常规生化检测项目性能水平.方法 选取2021年4月室间质量评价(EQA)回报结果中偏倚绝对值的平均值作为项目的偏倚估计,选取3个月室内质控累积在控变异系数作为不精密度估计,采用卫生行业标准WS/T 403-2012的允许总误差作为项目质量规范,绘制标准化西格玛性能验证图.对σ<6的...

关 键 词:西格玛  西格玛性能验证图  常规生化  性能评价  允许总误差  变异系数  偏倚

Application of Standardized Sigma Performance Verification Chart in Performance Evaluation of Routine Biochemical Testing Projects
Authors:CHEN Zhen  CHEN Tao  HUANG Zhen  XIONG Wei
Institution:(Department of Clinical Laboratory,Jishou People's Hospital,Jishou Hu'nan 416000,China;Department of Clinical Laboratory,Xiangxi National Hospital of Traditional Chinese Medicine,Jishou Hu'nan 416000,China)
Abstract:Objective To apply the standardized sigma verification chart to evaluate the performance level of laboratory routine biochemical testing items.Methods The average value of the absolute value of bias in the report results of EQA in April 2021 was selected as the bias estimation of the project,and the cumulative in-control coefficient of variation of indoor quality control in 3 months was selected as the imprecision estimation.The allowable total error of health industry standard WS/T 403-2012 was adopted as the project quality standard,and the standardized sigma performance verification chart was drawn.For projects withσ<6,according to the quality goal index(QGI),determine the priority improvement direction.Results UA,AST,ALP,α-HBDH and CHE reached the world-class level(σ≥6),ALT was excellent(5≤σ<6),TG and TBIL were good(4≤σ<5),Na,CHO,CK andγ-GT were critical(3≤σ<4),and the detection levels of K,CL,CREA,CA,P,GLU,UREA,TP,ALB,HDL-C,AMY,LDH and Fe were poor or unacceptable(σ<3).According to GQI value,5 projects need to improve precision first,11 projects need to improve bias first,and 4 projects need to improve precision and bias at the same time.Conclusion Standardized sigma performance verification chart can show theσlevel of each test item,which is convenient for laboratories to know the performance level of test items correctly and objectively.
Keywords:Sigma  Sigma performance verification chart  routine biochemistry  performance evaluation  allowable total error  coefficient of variation  bias