Non-contrast-enhanced MR angiography of the thoracic aorta using cardiac and navigator-gated magnetization-prepared three-dimensional steady-state free precession |
| |
Authors: | Amano Yasuo Takahama Katsuya Kumita Shinichiro |
| |
Institution: | Department of Radiology, Nippon Medical School, Tokyo, Japan. yas-amano@nifty.com |
| |
Abstract: | PURPOSE: To assess the usefulness of non-contrast-enhanced MR angiography using cardiac and navigator-gated magnetization-prepared three-dimensional (3D) steady-state free precession (SSFP) imaging for the diagnosis of diseases of the thoracic aorta. MATERIALS AND METHODS: Twenty-two patients with diseases of the thoracic aorta were examined using a 1.5 Tesla unit. Non-contrast-enhanced MR angiography was done using parasagittal 3D SSFP combined with cardiac-gating and k-space weighted navigator-gating techniques, using T2-prepared and fat-suppression pulses. Imaging quality and the diagnostic capability of this technique were compared with the imaging quality of 2D SSFP or contrast-enhanced 3D MR angiography and with final diagnoses. RESULTS: Non-contrast-enhanced 3D MR angiography provided signal-to-noise and contrast-to-noise ratios of the thoracic aorta comparable to non-contrast-enhanced 2D or contrast-enhanced 3D MR angiography (P > 0.17). This imaging technique gave accurate diagnoses in 19 of the 22 patients. CONCLUSION: Non-contrast-enhanced MR angiography using cardiac and navigator-gated magnetization-prepared 3D SSFP technique was useful for the diagnosis of diseases of the thoracic aorta. |
| |
Keywords: | MR angiography thoracic aorta steady‐state free precession navigator fat suppression |
本文献已被 PubMed 等数据库收录! |