Halogen and LED light curing of composite: temperature increase and Knoop hardness |
| |
Authors: | L. F. Schneider S. Consani L. Correr-Sobrinho A. B. Correr M. A. Sinhoreti |
| |
Affiliation: | (1) área Materiais Dentários, Departamento de Odontologia Restauradora, Faculdade de Odontologia de Piracicaba, Universidade Estadual de Campinas, Av. Limeira 901, Caixa Postal 52, CEP-13414-903 Piracicaba, Sao Paulo, Brazil |
| |
Abstract: | This study assessed the Knoop hardness and temperature increase provided by three light curing units when using (1) the manufacturers’ recommended times of photo-activation and (2) standardizing total energy density. One halogen – XL2500 (3M/ESPE) – and two light-emitting diode (LED) curing units – Freelight (3M/ESPE) and Ultrablue IS (DMC) – were used. A type-K thermocouple registered the temperature change produced by the composite photo-activation in a mold. Twenty-four hours after the photo-activation procedures, the composite specimens were submitted to a hardness test. Both temperature increase and hardness data were submitted to ANOVA and Tukey’s test (5% significance). Using the first set of photo-activation conditions, the halogen unit produced a statistically higher temperature increase than did both LED units, and the Freelight LED resulted in a lower hardness than did the other curing units. When applying the second set of photo-activation conditions, the two LED units produced statistically greater temperature increase than did the halogen unit, whereas there were no statistical differences in hardness among the curing units. |
| |
Keywords: | Light curing unit Hardness Irradiance Polymerization Temperature increase |
本文献已被 PubMed SpringerLink 等数据库收录! |
|