臭氧熏蒸对医院检验单消毒效果的试验观察
引用本文:朱桂华,章光华,钱宏彬,施玉国. 臭氧熏蒸对医院检验单消毒效果的试验观察[J]. 中国消毒学杂志, 1999, 16(2): 107-109
作者姓名:朱桂华  章光华  钱宏彬  施玉国
作者单位:安徽省天长市人民医院天长 239300
摘    要:经检测,臭氧对位于深层检验单表面细菌的杀灭作用较差,但当检验单上污染细菌总数为7~14cfu/cm2(包括真菌),并按≤3张一叠放于消毒器中,在温度为24℃,相对湿度为88%的条件下熏蒸消毒30min(臭氧浓度达0434mg/L),即未再检出细菌。

关 键 词:臭氧  医疗检验单消毒  细菌污染检测
修稿时间:1998-06-18

EXPERIMENTAL OBSERVATION ON EFFICACY OF OZONE FUMIGATION IN DISINFECTION OF LABORATORY REPORT SHEET IN HOSPITAL
Zhu Guihua,et al.. EXPERIMENTAL OBSERVATION ON EFFICACY OF OZONE FUMIGATION IN DISINFECTION OF LABORATORY REPORT SHEET IN HOSPITAL[J]. Chinese Journal of Disinfection, 1999, 16(2): 107-109
Authors:Zhu Guihua  et al.
Affiliation:Zhu Guihua,et al. Tianchang Municipal People's Hospital,Anhui Province,Tianchang 239300
Abstract:Examination indicated that the efficacy of ozone in killing the bacterial on surface of the report sheet located at the deep layer was low, but when the total number of bacteria contaminating the report sheets was 714 cfu/cm2 (including fungi) and the piles consisting of 3 or less than 3 report sheets were disinfected in disinfector by fumigation at a temperature of 24 and relative humidity of 85% for 30 min (the ozone concentration attained 0434 mg/L), bacteria were no longer detected.
Keywords:ozonedisinfection of medical laboratory report sheetexamination of bacterial contamination  