不同脉冲高强度聚焦超声辐照方式损伤离体灌注猪肝的实验研究
引用本文:焦娇,伍烽,邹建中,李发琪,刘芳,赵欣,王琦. 不同脉冲高强度聚焦超声辐照方式损伤离体灌注猪肝的实验研究[J]. 南方医科大学学报, 2013, 33(2): 230-234
作者姓名:焦娇  伍烽  邹建中  李发琪  刘芳  赵欣  王琦
作者单位:重庆医科大学生物医学工程学院,省部共建超声医学工程国家重点实验室,超声医学工程重庆市市级重点实验室,重庆400016
基金项目:国家重点基础研究发展计划973计划,国家自然科学基金重点项目,国家自然科学基金面上项目
摘    要:目的比较不同脉冲高强度聚焦超声(HIFU)辐照方式损伤离体灌注猪肝靶区临近含较大血管时的损伤情况,为临床应用
提供依据。方法建立离体猪肝经肝动脉和门静脉的体外循环灌注模型,在B超引导下选择直径为4 mm血管,于血管方向离管
壁3 mm位置给予相同剂量PHIFU辐照。实验组选择脉冲波,对照组用连续波,辐照结束后观察声通道血管及靶区损伤情况,
并取血管与靶区处组织送病理检查。结果PHIFU相同剂量辐照声通道内含4 mm大血管后,实验组和对照组的声通道血管均
未见损伤,两组凝固性坏死体积比较存在差异(P<0.05),实验组坏死体积大于对照组。声通道血管进行HE染色、血管弹力纤维
和胶原纤维染色显示没有损伤,实验组和对照组没有差别(P>0.05)。结论当靶区临近含有较大血管时,脉冲辐照能够提高
HIFU治疗效率,有效的形成损伤,且不会损伤血管。


关 键 词:脉冲高强度聚焦超声  离体猪肝  灌注模型  血管

Effect of ablations by pulsed versus continuous high-intensity focused ultrasound on isolated perfused porcine liver
JIAO Jiao , WU Feng , ZOU Jianzhong , LI Faqi , LIU Fang , ZHAO Xin , WANG Qi. Effect of ablations by pulsed versus continuous high-intensity focused ultrasound on isolated perfused porcine liver[J]. Journal of Southern Medical University, 2013, 33(2): 230-234
Authors:JIAO Jiao    WU Feng    ZOU Jianzhong    LI Faqi    LIU Fang    ZHAO Xin    WANG Qi
Affiliation:State Key Laboratory of Ultrasound Engineering in Medicine Co-founded by Chongqing and the Ministry of Science Technology,Chongqing key Laboratory of Ultrasound in Medicine and Engineering,College of Biomedical Engineering,Chongqing Medical University,Chongqing 400016,China
Abstract:Objective To compare the effect of ablations with pulsed and continuous high-intensity focused ultrasound (HIFU)
on the target area containing large blood vessels in isolated perfused porcine liver. Methods An isolated perfused liver model
was established by perfusing the isolated porcine liver with autologous blood through the portal vein and hepatic artery using
an cardiopulmonary bypass machine. With B-mode ultrasound guidance, a blood vessel about 4 mm in diameter was selected,
and pulsed and continuous HIFU ablation at the same dose was delivered along the blood vessel at 3 mm to the vascular wall.
After the ablation, the damages of the blood vessels as the acoustic channel and the target area were inspected, and the tissues
at the interface of target volume and blood vessels were sampled for pathological examination. Results Neither pulsed nor
continuous HIFU caused noticeable injuries of the acoustic channel vessels, but the volume of necrotic tissues in the target area
differed significantly between the two groups (P<0.05). HE staining and vessel staining did not show any injuries of the blood
vessels in the pathway of ultrasound ablation in neither of the groups (P>0.05). Conclusions When the target area contains
large blood vessels in the pathway of ultrasound ablation, pulsed HIFU can have better therapeutic efficiency to effectively
induce tissue damages without affecting the blood vessels.