胼胝体磁共振弥散张量成像技术参数研究
引用本文:方美芳,王震寰,沈龙山,单连强.胼胝体磁共振弥散张量成像技术参数研究[J].蚌埠医学院学报,2014,39(1):34-37.
作者姓名:方美芳  王震寰  沈龙山  单连强
作者单位:方美芳(蚌埠医学院第一附属医院,放疗科,安徽,蚌埠,233004,蚌埠医学院);王震寰(蚌埠医学院临床应用解剖研究所,人体解剖学教研室组织移植安徽省重点实验室,安徽,蚌埠,233030蚌埠医学院组织移植安徽省重点实验室,安徽,蚌埠,233030,);沈龙山(蚌埠医学院第二附属医院,放射科,安徽,蚌埠,233040);单连强(蚌埠医学院第二附属医院,放射科,安徽,蚌埠,233040);
摘    要:目的:探讨胼胝体磁共振弥散张量成像(DTI)临床应用的扫描参数.方法:对30名健康成年人行头部矢状位DTI,比较不同b值、层厚/层间距、扩散敏感梯度方向3个扫描参数对各向异性分数(FA)图像质量的影响.结果:不同的扫描参数对FA图像质量的影响各不相同.中间位b值(b=1与000 s/mm2)时FA 图像质量最佳,高b值(b=1 500 s/mm2) FA图像噪声大,低b值(b=500 s/mm2)FA图像对比度差.层厚/层间距越大,FA图像信噪比越高,但会遗漏细微结构的显示.13个与25个扩散敏感梯度方向FA图像质量差异无统计学意义(P〉0.05),6个扩散敏感梯度方向FA图像质量较差.结论:临床实际应用中,胼胝体矢状位DTI的适用扫描参数为b=1,000 s/mm2、5 mm/0 mm层厚/层间距、13个扩散敏感梯度方向.

关 键 词:  胼胝体  弥散张量成像  磁共振成像  技术参数

The parameters of magnetic resonance diffusion tensor imaging in corpus callosum
FANG Mei-fang,WANG Zhen-huan,SHEN Long-sha,SHAN Lian-qiang.The parameters of magnetic resonance diffusion tensor imaging in corpus callosum[J].Journal of Bengbu Medical College,2014,39(1):34-37.
Authors:FANG Mei-fang  WANG Zhen-huan  SHEN Long-sha  SHAN Lian-qiang
Institution:1. Department of Radictheraphy, The First Affiliated Hospital of Bengbu Medical College, Bengbu Anhui 233004; 2. Institution of Clinic Anatomy, Department of Human Anatomy, 3. Anhui Key Laboratory of Tissue Transplantation,Bengbu Medical College,Bengbu Anhui 233030; 4. Department of Radiology, The Second Affiliated Hospital of Bengbn Medical College, Bengbu Anhui 233040, China)
Abstract:Objective:To explore the technical parameters of magnetic resonance diffusion tensor imaging(DTI) in corpus callosum. Methods:The DTI of the sagittal plane were detected in 30 nomarl brains. The effects of b value, slice thickness/interlamellar spacing and diffusion sensitive gradient direction on fractional anisotropy(FA) image quality were compared. Results:Different parameters had different effects on FA image quality. The quality of FA image was best when b value was middle( b = 1 000 s/mm2) ,FA image noise was large at the high b value(b = 1 500 s/mm^2 ) ,and the contrast of FA image was poor at the low b value(b =500 s/mm2 ). The signal to noise ratio was high with slice thickness/interlamellar spacing increasing, the subtle structure was easy to miss. The differences of FA image qualities at 13 and 25 diffusion sensitive gradient directions statistical significance ( P 〉 0.05 ). The FA image quality at 6 diffusion sensitive gradient direction was worse hadno. Conclusions:The 1 000 s/mm2 of b value,5 mm/0 mm of slice thickness and 15 diffusion sensitive gradient directions are suitable for DTI of corpus callosum in the sagittal head in clinic.
Keywords:brain  corpus callosum  diffusion tensor imaging  magnetic resonance imaging  technical parameter