首页 | 本学科首页   官方微博 | 高级检索  
     


Complete dislocation of the polyethylene inlay after internal hemipelvectomy and individually fitted tumor endoprosthesis
Authors:Lichtinger T K  Lederer M  Müller R T
Affiliation:Orthop?dische Universit?tsklinik Essen.
Abstract:We report about an exceptional complication after internal hemipelvectomy and replacement of the defect with a custom-made endoprosthesis. A complete luxation of the PE-inlay out of the metal cup occurred. Radiologically we assumed a luxation of the femoral head in dorsal direction. Revision after a failed attempt of closed reposition showed a complete luxation of the inlay, which was caused by deficient fixation in the acetabular component. The absence of a contrast wire in the PE-inlay delayed the right diagnosis and made it difficult to find the dislocated inlay in the large wound.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏