The accuracy of transoesophageal echocardiography in estimating pulmonary capillary wedge pressure in anaesthetised patients |
| |
Authors: | Ali M M Royse A G Connelly K Royse C F |
| |
Affiliation: | Anaesthesia and Pain Management Unit, Department of Pharmacology, The University of Melbourne, Australia. |
| |
Abstract: | The objective of this study was to identify whether pulmonary capillary wedge pressure can be estimated in anaesthetised patients receiving mechanical ventilation, using transoesophageal echocardiography. A retrospective validation study investigated a 10-patient cohort with variable haemodynamic conditions, and a 102-patient series in which a single measurement was made during stable haemodynamic conditions. Concurrent echocardiographic Doppler and pulmonary artery catheter wedge pressure measurements were performed. In the 10-patient cohort, the systolic fraction of Doppler measurements in the pulmonary vein (r = -0.32, p = 0.035) and the E/A ratio (r = 0.56, p = 0.0009) were correlated with the wedge pressure. In all cases, the limits of agreement exceeded 10 mmHg, and sensitivity or specificity for detecting wedge pressure ≥ 15 mmHg was poor. This study demonstrates proof of concept that using transoesophageal echocardiography for estimating the pulmonary artery wedge pressure may not be sufficiently accurate for clinical use. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|