无体模法定量CT测量腰椎骨密度的可信度及可重复性研究
引用本文:张望,彭松林,镇万新,罗常.无体模法定量CT测量腰椎骨密度的可信度及可重复性研究[J].骨科,2016,7(6):390-393.
作者姓名:张望  彭松林  镇万新  罗常
作者单位:暨南大学第二临床医学院 深圳市人民医院脊柱外科, 广东深圳,518020
基金项目:深圳市科技计划项目(GJHZ20140416123146582)
摘    要:目的 评价无体模法定量CT(quantitative computed tomography,QCT)测量腰椎骨密度(bone mineral density,BMD;QCT-BMD)的同一位测量者内和3位测量者间BMD测量值的一致性和可重复性.方法 将31例患者的腰椎螺旋CT扫描数据传输至工作站进行BMD测量.由3位经过培训的脊柱外科医生(分别为测量者A、B、C)独立进行腰椎QCT-BMD测量,每位测量者对患者均测量2次.同一位测量者的2次测量和3位测量者间BMD测量值的差异用重复测量的方差分析比较,一致性分析用组内相关系数(intraclass correlation coefficient,ICC)描述.3位测量者间BMD测量值的可靠性分析用Cron-bach'sα系数描述.结果 测量者A的2次腰椎BMD测量值分别为(67.14±26.71)mg/cc、(66.37±26.69)mg/cc,差值率为4.22%±2.55%(P=0.12,ICC=0.995);测量者B的2次腰椎BMD测量值分别为(68.89±29.37)mg/cc、(68.54±28.98)mg/cc,差值率为5.08%±3.83%(P=0.56,ICC=0.994);测量者C的2次腰椎BMD测量值分别为(69.31±26.72)mg/cc、(69.83±27.90)mg/cc,差值率为4.16%±3.03%(P=0.31,ICC=0.995).3位测量者所测量的腰椎BMD平均值分别为(66.76±26.66)mg/cc、(68.71±29.13)mg/cc、(69.57±27.28)mg/cc(P=0.15,ICC=0.957,Cronbach'sα系数=0.985).结论 无体模法QCT测量腰椎BMD具有较高的一致性和可重复性.

关 键 词:骨密度  定量CT  可信度  可重复性
收稿时间:2016/4/11 0:00:00

Reproducibility of measurement of lumbar vertabal bone mineral density using quantitative CT without external reference phantom
ZHANG Wang,PENG Songlin,ZHEN Wanxin and LUO Chang.Reproducibility of measurement of lumbar vertabal bone mineral density using quantitative CT without external reference phantom[J].Orthopaedics,2016,7(6):390-393.
Authors:ZHANG Wang  PENG Songlin  ZHEN Wanxin and LUO Chang
Abstract:Objective To evaluate the intra- and inter-observer consistency and reproducibility of bone mineral density (BMD) in lumbar vertebrae measured by quantitative CT without external reference phantom. Methods The CT data of 31 cases were transferred to a work station to measure lumbar BMDs. The BMDs were measured by three trained spine surgeons. Each observer performed the measurements of all subjects twice. Intra-observer and inter-observer variability was calculated using analysis of variance of the repeatedly measured results. The consistency was analyzed with the intra-class correlation coefficient (ICC). Reliability analysis in inter-observer was performed with Cronbach's α coefficient. Results Data measured by Observer A were (67.14 ± 26.71) mg/cc and (66.37 ± 26.69) mg/ccdifference: (4.22 ± 2.55) %, P=0.12, ICC=0.995], Observer B (68.89 ± 29.37) mg/cc and (68.54 ± 28.98) mg/ccdifference:(5.08 ± 3.83)%, P=0.56, ICC=0.994], Observer C (69.31 ± 26.72) mg/cc and (69.83 ± 27.90) mg/ccdifference:(4.16 ± 3.03)%, P=0.31, ICC=0.995], respectively. The mean lumbar BMDs by the three observers were (66.76 ± 26.66), (68.71 ± 29.13) and (69.57 ± 27.28) mg/cc, respectively (P=0.15, ICC=0.957, Cronbach's α coefficient=0.985). Conclusion High reproducibility and consistency in the measurement of lumbar BMD can be achieved with quantitative CT without external reference phantom.
Keywords:Bone mineral density  Quantitative CT  Reliability  Reproducibility