An under-occluded roller pump is less hemolytic than a centrifugal pump
Authors:Rawn D J  Harris H K  Riley J B  Yoda D N  Blackwell M M
Institution:Clinical Services Department, College of Health Professions, Medical University of South Carolina, Charleston, USA.
Abstract:The purpose of this study is to measure and compare the hemolysis produced by roller pumps with varied occlusion settings and a centrifugal pump. The null hypothesis is that there is no difference in the Index of Hemolysis (IH = gm Hb/100 L pumped) produced by a roller pump (RP) at four different occlusion settings and a centrifugal pump (CP) at the same blood flow rate (4.5 L/min) and afterload (250 mmHg, +/-10 mmHg) over three hours. Five identical closed-loop circuits were assembled and primed with saline. The pumps were then calibrated and occlusions were set. In three circuits, the occlusion for the RP was opened at 5 RPMs to support 150, 225, or 300 mmHg (+/-10 mmHg) against a clamped line. In one circuit, a RP was adjusted to a barely non-occlusive setting (1 cm drop/30 inch gradient). The fifth circuit employed a CP. Prior to testing, the saline in each circuit was replaced with one liter of fresh bovine blood (Hct = 22 +/- 2%). The IH for each treatment was compared in six trials yielding a statistical power > 0.80. Analysis of variance with multiple comparison (p < or = 0.05) demonstrated that compared to the barely non-occlusive setting, the IH in the centrifugal pump was not significantly greater. Under-occluded RP settings yielded IHs significantly less than the CP. It appears that opening the occlusion on a roller pump allows a lower IH compared to traditional RP occlusion setting or centrifugal pumping.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号