Correlation of Noninvasive Blood Pressure and Invasive Intra-arterial Blood Pressure in Patients Treated with Vasoactive Medications in a Neurocritical Care Unit
Authors:Ali A. Saherwala  Sonja E. Stutzman  Mohamed Osman  Junaid Kalia  Stephen A. Figueroa  DaiWai M. Olson  Venkatesh Aiyagari
Affiliation:1.Division of Neurocritical Care, Department of Neurology and Neurotherapeutics,UT Southwestern Medical Center,Dallas,USA;2.Department of Critical Care,Aurora St. Luke’s Hospital,Milwaukee,USA;3.Department of Neurological Surgery,UT Southwestern Medical Center,Dallas,USA
Abstract:

Background

The correlation between noninvasive (oscillometric) blood pressure (NBP) and intra-arterial blood pressure (IAP) in critically ill patients receiving vasoactive medications in a Neurocritical Care Unit has not been systematically studied. The purpose of this study is to examine the relationship between simultaneously measured NBP and IAP recordings in these patients.

Methods

Prospective observational study of patients (N = 70) admitted to a neurocritical care unit receiving continuous vasopressor or antihypertensive infusions. Paired NBP/IAP observations along with covariate and demographic data were abstracted via chart audit. Analysis was performed using SAS v9.4.

Results

A total of 2177 paired NBP/IAP observations from 70 subjects (49% male, 63% white, mean age 59 years) receiving vasopressors (n = 21) or antihypertensive agents (n = 49) were collected. Paired t test analysis showed significant differences between NBP versus IAP readings: ([systolic blood pressure (SBP): mean = 136 vs. 140 mmHg; p < 0.0001], [diastolic blood pressure (DBP): mean = 70 vs. 68 mmHg, p < 0.0001], [mean arterial blood pressure (MAP): mean = 86 vs. 90 mmHg, p < 0.0001]). Bland–Altman plots for MAP, SBP, and DBP demonstrate good inter-method agreement between paired measures (excluding outliers) and demonstrate NBP–IAP SBP differences at extremes of blood pressures. Pearson correlation coefficients show strong positive correlations for paired MAP (r = 0.82), SBP (r = 0.84), and DBP (r = 0.73) recordings. An absolute NBP–IAP SBP difference of > 20 mmHg was seen in ~ 20% of observations of nicardipine, ~ 25% of observations of norepinephrine, and ~ 35% of observations of phenylephrine. For MAP, the corresponding numbers were ~ 10, 15, and 25% for nicardipine, norepinephrine, and phenylephrine, respectively.

Conclusion

Despite overall strong positive correlations between paired NBP and IAP readings of MAP and SBP, clinically relevant differences in blood pressure are frequent. When treating with vasoactive infusions targeted to a specific BP goal, it is important to keep in mind that NBP and IAP values are not interchangeable.
Keywords: